Technical Information
| Driftcon system components |
|
The DRIFTCON system consists of the following components:
- DRIFTCON-16 hardware module
- Intelligent CyclerProbe Fixture (available in several models and formats)
- 16 x 0.2ml: 16 temperature sensors for 0.2ml block format
- 16 x 0.5ml: 16 temperature sensors for 0.5ml block format
- 16 x 384 fit: 16 temperature sensors for 384 block format
- Other special models available on request
- DRIFTCON SMART-card (10 acquisitions)
- Connection Cable
- DRIFTCON Software
- Robust, CyclerProbe Fixture box
- Optional: State-of the-art Notebook with DRIFTCON software pre-installed
|
|
| System features |
- Dynamic Temperature Monitoring (DTM)
- Static Temperature Verifications (STV)
- Multi-channel (up to 16)
- Intelligent temperature probes fixtures
- NIST-traceable, validated by Dutch Council for Accreditation
- Main supply 90-240VAC (50-60Hz)
- CE-mark
|
|
| System Specifications |
|
Specifications are subject to change without prior notice
- Accuracy1: average channel value accuracy up to <0.1°C or better (NIST-traceable)
- Measurement resolution: 16bits
- Software sample rate/channel: 2Hz
- Measurement data range: 10 – 100°C
(1. Temperature range 30.0 -95.0°C)
|
|
| Driftcon Applications |
- GMP/GLP validations
- FDA regulations, validations
- Diagnostic quality programs
- Other quality program qualifications
- Protocol transfer and cycler transfer
- Instrument to instrument comparison
- Periodic performance validations
- Process problem solving
- Real-time PCR Instrument validation
- Biotechnology applications, Clinical laboratory QC
- PCR® plate temperature uniformity
|
|
|
|